The system for the monitoring of the photomultiplier dark noise in the Borexino neutrino detector

D.I. Orekhov$^1$, A.S. Chepurnov$^1$, A.A. Sabel’nikov$^2$, A.V. Etenko$^2$

Moscow University Physics Bulletin 2008. 63. N 3. P. 203

  • Article
Annotation

Monitoring the dark noise of photomultiplier tubes (PMT) is important for maintaining the operation of the Borexino detector. For this purpose, a dark noise monitoring system was developed on the basis of the CAN network and the DeviceNet upper-level protocol. The system consists of a set of CAN controllers, which digitize PMT signals and transmit data over the CAN to the host computer. The software creates a DeviceNet protocol stack, ensuring data collection and transmission. The user interface is built on the basis of dynamic WEB pages.