Annotation
In this work the application of a rather simple half-analytic model for plotting and studying electron trajectories in an aqueous environment in a transverse magnetic field was considered. The energies of electrons used in medical electron accelerators, viz., those up to 30 Me, were considered and the induction value of the magnetic field was changed within limits up to 3 T. Our data are compared with those in the works of other authors, who used the Monte Carlo simulation method, as well as with experimental data.
Received: 2010 October 3
Approved: 2011 November 11
PACS:
34.10.+x General theories and models of atomic and molecular collisions and interactions
© 2016 Publisher M.V.Lomonosov Moscow State University
Authors
A.V. Belousov, A.S. Namestnikov, A.P. Chernyaev
Department of Physics of High Energy Accelerators, Faculty of Physics, Moscow State University, Moscow, 119991, Russia
Department of Physics of High Energy Accelerators, Faculty of Physics, Moscow State University, Moscow, 119991, Russia